Best Maintenance Practices for How to Maintain Portable Suction Units

Written by Sam D. Say | Feb 12, 2018, 12:30:00 PM

A portable suction unit is a medical device designed to remove fluids, secretions, and foreign materials from a patient's airway to ensure patency. These devices utilize negative pressure to aspirate blood, vomit, or mucus through a catheter into a collection canister. Proper maintenance of these units is vital for effective airway management and preventing hypoxia during emergency medical response. Learning how to maintain portable suction units is a critical skill for any medical professional, as it ensures the device is ready for immediate use.

 

Key Takeaways

    • Portable suction units use negative pressure to clear patient airways and prevent life-threatening hypoxia incidents.
    • Regular maintenance of suction devices ensures they are functional and ready for critical emergency medical scenarios.
    • Battery health checks and power reliability tests are essential components of routine suction unit maintenance protocols.
    • Strict decontamination procedures prevent pathogen transmission by requiring the disposal of all used suction unit components.
    • Proper storage environments protect portable suction equipment from extreme temperatures, moisture, and potential mechanical debris damage

 

Clinical Applications of Emergency Portable Suction

While portable suction units are often associated with respiratory emergencies, they are essential tools for managing various clinical emergencies, including:

  • Stroke
  • CHF
  • Inability to swallow
  • Altered consciousness
  • Overdose/intoxication
  • Foreign body obstruction
  • Trauma

 

Any emergency can turn into a respiratory emergency, and your primary responsibility is to ensure a patent airway and prevent hypoxia. Some of the signs of hypoxia include:

  • Increased work of breathing
  • Use of accessory muscles
    • Belly-breathing (in children), nasal flaring, retractions
  • Restlessness/agitation
  • Confusion
  • Bradycardia, especially in pediatrics
  • Decreased level of consciousness
  • Inadequate breath sounds
  • Abnormal breath sounds (gurgling, wheezing, stridor)
  • Coma
  • Seizures

The critical role of suction in preventing hypoxia means you must ensure your unit is functional. This comes down to proper maintenance. Understanding the nuances of how to maintain portable suction units prevents equipment failure during these high-stakes interventions.

Ensuring Power Reliability and Battery Health for Suction Devices

Your portable suction unit is of no use if it doesn’t power up. Most units rely on rechargeable batteries, but as we all know, batteries can fail, which is why it is crucial to check your suction unit at the start of each shift. If your unit fails to power up, here are a few tasks to check before doing anything else:

    • Ensure the canister is properly connected, all ports on the canister are capped and the patient tubing is firmly attached.
    • If the regulator is at a very low setting, adjust the regulator.
    • Replace the canister if the filter becomes clogged.
    • Replace the canister if full.

If these mechanical factors are not the cause, the issue may stem from battery degradation. It is critical to check the overall health of your batteries—not just at the start of each shift, but as part of your monthly or quarterly truck maintenance. Here are our guidelines:

    • Confirm the power cord is supplying power to the device and check the power indicator light on the control panel.
    • Remove the power cord.
    • Turn the device on and check for vacuum by occluding the canister.
    • Allow the unit to run for 15 minutes on DC power. If the unit stops or slows during the 15 minutes, or if the battery condition indicators begin to blink, it is possible the battery capacity has been depleted and it is time to replace the battery.
    • If the unit is still running at full power after 15 minutes, turn the device off and put it back on charge.
    • Reconnect the device to the charging source.

Decontamination Protocols for Preventing Pathogen Transmission

Maintaining cleanliness is a critical aspect of suction unit maintenance because these devices are inherently prone to pathogen contamination, as they come in contact with all forms of bodily fluids (vomit, sputum, blood). Don’t skimp when decontaminating your unit! Here are our recommendations:

    • Always wear personal protective equipment when handling contaminated equipment:
      • Gloves
      • Face and eye protection
      • Protective clothing
  • Disconnect the unit from any power source prior to cleaning.
  • Sanitize after each use, even if nothing shows in the canister.
    •  Use a mild detergent or a mixture of bleach and water (one-part bleach/10-parts water) and rinse thoroughly.
  • Disinfect all outer surfaces of the unit, including control knobs, screens, and handles.
  • Discard ALL disposable parts: canister, patient tubing, and catheters.
  • NEVER reuse disposable parts. They are disposable for a reason.
  • Follow appropriate guidelines for disposal of biohazardous materials.
  • Follow manufacturer’s instructions when disinfecting the mechanics of the unit.
  • Do not submerge your suction unit.
  • Disconnect the battery from the PC board when cleaning the interior chassis.

Storage and Environmental Protection for Portable Suction Equipment

The storage environment of a portable suction unit is a critical factor in its long-term reliability and maintenance. Here are a few guidelines:

    • Do not expose the unit to extreme temperatures.
    • Avoid wet or high moisture environments.
    • Secure the unit to prevent damage.
    • Avoid areas of dust or debris.

By following these best maintenance practices for your portable suction unit, you can be confident your equipment will be ready and able to handle the next suction scenario. Consistency in how to maintain portable suction units will ultimately improve patient outcomes in emergency settings.

 